WEIGHT
2.8oz / 80g ( main unit)
DEVICE DIMENSIONS
7.00'' x 4.84'' x 2.95'' (main unit)
DISPLAY
2.7'' touch screen, HD
CONNECTOR
Micro D connector
WIRELESS CONNECTIVITY
Built-in Bluetooth dual mode, support 4.0 BLE
CHARGE TIME
Less than 2 hours to 90%
BATTERY TYPE
Rechargeable lithium-polymer battery
LEAD TYPE
Integrated EKG electrodes
External EKG cable and electrodes
LEAD SET
Lead I, lead II
DISPLAY GAIN
1.25 mm/mV, 2.5 mm/mV, 5 mm/mV 10 mm/mV, 20 mm/mV
SAMPLING RATE
500Hz
SAMPLING ACCURACY
16 bit
MEASUREMENT MODE
Episode, continuous
SWEEP SPEED
25 mm/s
ACCURACY
±2 bpm or ±2%, whichever is greater
HR MEASUREMENT RANGE
30 to 250 bpm
ST MEASUREMENT RANGE
-0.5 to +0.5 mV
BAND WIDTH
0.05 to 40 Hz
ELECTRODE OFFSET POTENTIAL TOLERANCE
±300 mV
MEASUREMENT SUMMARY
Heart rate, QRS duration, ST segment, QT/QTc Rhythm analysis (Regular EKG Rhythm, High Heart Rate, Low Heart Rate, High QRS Value, High ST Value, Low ST Value, Irregular EKG Rhythm,Unable to analyze)
TYPE STANDARD
Standards
SpO2 RANGE
70% to 100%
MEASUREMENT MODE
Episode, continuous
SpO2 ACCURACY (ARMS)
80-100%: ±2%, 70-79%: ±3%
PR RANGE
30 to 250 bpm
PR ACCURACY
±2 bpm or ±2%,whichever is greater
PI RANGE
0.5-15
MEASUREMENT SUMMARY
SpO2, PR, PI, Summary (Normal Blood Oxygen, Low Blood Oxygen, Unable to analyze)
WEIGHT
5.6oz /160g
DEVICE DIMENSIONS
1.5'' x 1.2'' x 1.5'' (68*25mm, main unit)
CUFF SIZE
9'' - 17''/22-42cm
WIRELESS
Bluetooth 4.0 BLE
RECORDED PARAMETERS
Diastolic blood pressure, systolic blood pressure, pulse rate, mean arterial pressure, pulse pressure.
MEASURE TECHNOLOGY
Oscillometric
MethodCUFF PRESSURE RANGE
0-300 mmHg
BLOOD PRESSURE ACCURACY
±3 mmHg
PULSE RATE RANGE
30 to 250 bpm
PULSE RATE ACCURACY
±2 bpm
DEGREE OF DUST & WATER RESISTANCE
IP22
BATTERY LIFE
Rechargeable lithium-polymer, DC 5V. One charge for 1000 measurements.
iOS COMPATIBLE
iOS 9.0 or above. iPhone 4s or above; iPad 3 or above.
ANDROID COMPATIBLE
Android 5.0 or above, with Bluetooth 4.0 BLE.
MONITORING TIME
Up to 10 hours
DATA STORAGE
Store SpO2 and pulse rate
MEASUREMENT SUMMARY
Total duration, <90% STAT, Average saturation, Lowest saturation, Summary (No abnormal detected, blood oxygen drop detected, Unable to analyze)
MEASUREMENT RANGE
94-108°F
MEASUREMENT RANGE
20 to 600mg/dL (1.1 to 33.3 mmol/L)
Use left/right arrows to navigate the slideshow or swipe left/right if using a mobile device